INSTALLATION INSTRUCTIONS

CSMRI_F-3732_050813

MRI INSTALLATION REGISTRATION FORM

For warranty purposes, please fill out this form and return to Kenall by fax at (847) 360-1781.

I certify that the lighting installation for the listed MRI suite location is completed per the provided 

installation instructions and to the best of my abilities. 

Please check off items to denote status:

 

Installation instruction sheets for MRIPS-312 power supply and individual luminaire(s) read 

 

and followed.

 

MRIPS-312 power supply and EMI fi lters are located outside the shielded enclosure.

 

All DC supply wiring is completely enclosed within grounded aluminum conduit. Installation

  

has no ungrounded/unshielded portions of conduit or openings of any size or shape.

 

All dimming signal wiring is completely enclosed within grounded aluminum conduit. 

 

Installation has no ungrounded/unshielded portions of conduit or openings of any size or

 

shape. Check here if dimming is not applicable: 

 

MRI Room EMI fi lters for the 24VDC supply are of the type intended for MRI suites and are  

 

 

sized to the electrical load.

 

DC supply power and dimming signal are NOT running through the same EMI fi lter. Check here   

 

if dimming is not applicable: 

 

 

Lighting system fully tested (including dimming operation, if applicable) while MRI machine is    

 

in idle and scan operation mode.

If any of these steps cannot be completed or you need technical assistance, please contact Kenall 

Technical Support at 1-800-4KENALL (1-800-453-6255).

Electrical Contractor

Name: ___________________________________

City/State: ________________________________

Phone:___________________________________

Installation Date: ___________________________

Installation Site

Name: ___________________________________

City/State: ________________________________

FAX FORM TO (847) 360-1781

(Do not write below line)

Kenall Received By: ___________________________________    Received Date: ________________________________

Summary of Contents for CSMRI Series

Page 1: ...OUNTING RESTRICTIONS SAVE THESE INSTRUCTIONS INSTALLATION INSTRUCTIONS PREP INSTALLATION SURFACE Disconnect power to all circuits before wiring system Install in accordance with all national state and...

Page 2: ...age is divided among multiple lightboxes make sure lightbox is oriented toproduce a continuous image It is recommended that the door remain attached to the lightbox housing during installation This wi...

Page 3: ...supply from the EMI filter to luminaires s using at least 18 AWG wires Class 1 wiring is required Make sure wiring is completely enclosed in grounded aluminum conduit Any gaps regardless of size must...

Page 4: ...MRIPS 312 instruction sheet and fax to Kenall at 847 360 1781 II Door Opening Closing and Removal Procedure NOTE OPENING THE DOOR DURING NORMAL INSTALLATION IS NOT NECESSARY ALL WIRING SHOULD BE ACCE...

Page 5: ...for proper engagement of catch hooks before closing door Installing the Door CSMRI22 Lift door by spring clips on one side Engage spring clips on one side to hang vertically Closing the Door With catc...

Page 6: ...on opposite sides of each other and when installed into the lightbox will put the image in the correct orientation Fasten all hex nuts to retainer brackets Torque all fasteners equally to 12 lb in Re...

Page 7: ...ed installed in applications for which it is not intended This includes operation in ambient temperatures above stated limits for any length of time Failure by electrical surge shall not be covered un...

Page 8: ...to an ungrounded supply Wiring connections must be made internal to the enclosure Failure to install unit on a mechanically sound surface may result in personal injury physical damage or potential fi...

Page 9: ...nduit Installation has no ungrounded unshielded portions of conduit or openings of any size or shape Check here if dimming is not applicable MRI Room EMI filters for the 24VDC supply are of the type i...

Reviews: