www. kenall.com P: 800-4-Kenall F: 847-360-1781 1020 Lakeside Drive Gurnee, Illinois 60031
9
INSTALLATION INSTRUCTIONS
CSMRI_F-3732_050813
MRI INSTALLATION REGISTRATION FORM
For warranty purposes, please fill out this form and return to Kenall by fax at (847) 360-1781.
I certify that the lighting installation for the listed MRI suite location is completed per the provided
installation instructions and to the best of my abilities.
Please check off items to denote status:
Installation instruction sheets for MRIPS-312 power supply and individual luminaire(s) read
and followed.
MRIPS-312 power supply and EMI fi lters are located outside the shielded enclosure.
All DC supply wiring is completely enclosed within grounded aluminum conduit. Installation
has no ungrounded/unshielded portions of conduit or openings of any size or shape.
All dimming signal wiring is completely enclosed within grounded aluminum conduit.
Installation has no ungrounded/unshielded portions of conduit or openings of any size or
shape. Check here if dimming is not applicable:
MRI Room EMI fi lters for the 24VDC supply are of the type intended for MRI suites and are
sized to the electrical load.
DC supply power and dimming signal are NOT running through the same EMI fi lter. Check here
if dimming is not applicable:
Lighting system fully tested (including dimming operation, if applicable) while MRI machine is
in idle and scan operation mode.
If any of these steps cannot be completed or you need technical assistance, please contact Kenall
Technical Support at 1-800-4KENALL (1-800-453-6255).
Electrical Contractor
Name: ___________________________________
City/State: ________________________________
Phone:___________________________________
Installation Date: ___________________________
Installation Site
Name: ___________________________________
City/State: ________________________________
FAX FORM TO (847) 360-1781
(Do not write below line)
Kenall Received By: ___________________________________ Received Date: ________________________________