display flashing C letter (do not blow into the
mouthpiece).
■
Next, the screen will display error notification
Er-b - press and hold TEST button.
■
The screen will display UnIt. Press and hold the
TEST button again. The screen will display ALAr.
■
To change alarm point, repeatedly press
TEST button; the device will display alarm
point value (available alarm points for
chosen units: % - 0,020, 0,030, 0,050,
0,080; mg/L – 100, 150, 250, 400; ‰ -
0,200, 0,300, 0,500, 0,800).
■
In order to save, leave the screen displaying
desired alarm point and wait until device
emits sound signal – change has been
successfully saved
Specifications
Electrochemical sensor (Fuel cell sensor)
4 digits display
LCD with backlight
Blowing time about 6 seconds
Blowing flow about 10 L/min
Blowing flow interrupt detection
Display range: 0.00~3.00‰(1.50 mg/L)
Accuracy: +/- 0.05‰ at 0.50‰
Warm-up time: 10 seconds
Testing time: 10 seconds
Repeat test: less than 60 seconds
Selective unit of measure: %, ‰, mg/L
Show „HI” if above the display range
LCD change color if above alcohol limit
Selective alarm point for 0.02~0.08% (default is 0.05%)
Auto power off
Test counter
Calibration interval: 1 year or 200 tests
Low battery indication
5 pcs mouthpiece included
Power supply: 1x AAA battery
Operation temp: 10ºC ~ 40ºC
Storage temp: 0ºC ~ 50ºC