Factory setting of the mixed mode function

When  delivered,  the  pyrometer  is  already  set  to  mixed-mode  operation 
and the parameters are configured according to the table. 

Configuration layer 

C004

 

Parameter 

PA 40 AF 1 

PA 40 AF 4 

PA 50 AF 1 

MIX.

 

ON 

ON 

ON 

RNG._

500 °C 

600 

400 °C 

RNG.~

1700 °C 

2400 °C 

1400 °C 

TNS._ 

640 °C 

740 °C 

490 °C 

TNS.~

660 °C 

760 °C 

510 °C 

SRC.1 

L2 

L2 

L2 

SRC.2 

SRC.C 

L2 

L2 

L2 

Configuration layer 

C010

 

Parameter 

PA 40 AF 1 

PA 40 AF 4 

PA 50 AF 1 

AOI.S 

MIX

 

MIX

 

MIX

 

AOI._

500 °C 

600 °C 

400 °C 

AOI.~

1700 °C 

2400 °C 

1400 °C 

Adjustment of the measured values 

In order to ensure a linear transition of the measured values, the adjust-
ment must be carried out as follows during commissioning: 

1. Align the pyrometer to the target object. Set the two-colour emissivi-

ty correction to 100 %.

2. Heat  up  the  object  to  a  temperature  approx.  in  the  middle  of  the

transition  range.  Wait  certain  times  until  the  temperature  is  stabili-
ses.

3. Make a note of the displayed quotient temperature.

4. Check  the  spectral  temperature  L1  or  L2.  (depending  on  the  set

source  1).  The  quotient  temperature  and  the  spectral  temperature
should  be  the  same.  If  the  spectral  temperature  is  higher  than  the
quotient temperature, increase the emissivity, otherwise decrease it.
The  emissivity  L1  can  be  set  in  code  page  "

C002

"  parameter

EPS1

 or L2 in code page "

C003

" parameter 

EPS2

.

Summary of Contents for PYRO CellaTemp PA

Page 1: ...PYROMETER CellaTemp PA Mixed Mode Function Ident no 1117971 06 2022 MANUAL...

Page 2: ...Operating manual...

Page 3: ...Mode 2 4 Factory setting of the mixed mode function 3 5 Adjustment of the measured values 3 6 Setting parameters at the pyrometer 4 6 1 Setting the parameters using the button 4 6 2 Operation of the...

Page 4: ...Operating manual...

Page 5: ...start and Measuring range end 2 Mixed mode function for extending the measuring range With mixed mode operation the measuring range can be extended to lower initial temperatures by linking a spectral...

Page 6: ...mbda 1 L2 Lambda 2 Q Quotient SRC C Selction of the con trolling source L1 Lambda 1 L2 Lambda 2 Q Quotient SAVE Save Save changes exit menu ESC Escape Discard changes exit menu To operate the pyromete...

Page 7: ...s In order to ensure a linear transition of the measured values the adjust ment must be carried out as follows during commissioning 1 Align the pyrometer to the target object Set the two colour emissi...

Page 8: ...VE 6 Now use MODE to either apply the changes you have made to the settings SAVE or to close without saving these changes END After that the display panel will resume showing the tem perature reading...

Page 9: ...Show this help site J Show diagnosis W Show ambient temperature Q Show calibration data X Show measure temperatures P Show channel parameters 6 2 2 Parameter diagnostic overview Command P shows an ov...

Page 10: ...ixed Mode you can adjust all relevant parameters In the sub menu Mixed Mode you can adjust all relevant parameters Activation deactivation of the temperature calculation Measuring range of the Mixed M...

Page 11: ...command T in the main menu you can select the source for the analogue output Select 4 for Mixed Mode The display shows the Mixed Mode temperature T Set Analog Out 1 SOURCE 1 Lambda 1 2 Lambda 2 3 Quot...

Page 12: ...or in part for any purpose whatsoever without prior consent of the author This applies to any form of mechanical or electronic reproduction as well as to electronic transmission in any form through a...

Reviews: