Keller K-114 BT Operating Instructions Manual Download Page 5

– 26 –

– 27 –

Product overview

Product

Connection

Product number

Plug compatible with

K-114 BT

Screw terminals 

(MC 1,5 / 6-ST-3,81)

309010.0119

all digital KELLER products with cable outlet, 

series 3X, series 4X, DCX

K-114 BT A

Fischer plug 

(S 103 A054-130 [male])

309010.0101  

DCX-16 /-22-25 PVDF / -38, LEO Record, 

LEX 1, GSM-2

K-114 BT B

Binder cable connector 

(series 680 [female] 5-pole)

309010.0120  

Series 30X / 40X, LEO 3, dV-22 PP, dV-2 PS*, 

Castello*

 *Requires additional cable option

Pin assignment for the converters and connections 

+Vcc RS GND RS

485-A

485-B

U-IN Erde /

Schirm

K-114 BT

Pluggable screw terminal

MC 1,5 / 6-ST-3,81 

K-114 BT A

Fischer connector

S 103 A054-130 (male)

K-114 BT B

Binder cable connector

Series 680 (female) 5-pole

3: +Vcc

4: RS485-A

5: RS485-B

2: +U-IN

1: GND

2: +U-IN

3: +Vcc

4: RS485-A

1: GND

5: RS485-B 

Earth/shield

Earth/shield

RS485 interface 

For Bluetooth communication and the internal controller (bus address: 253), the converter supports the baud rate of 

9600 baud (default) or 115200 baud, 8 data bits, 1 stop bit and no parity. This transmission speed must be configu-

red on the master (PC) and on the K-114 BT.

Property

Bluetooth / RS485

USB / internal controller (bus address 253)

configurable

Baud rate

9600 (default) / 115200 baud

9600 (default) / 115200 baud

Yes

Data bits

8

8

No

Stop bits

1

1

No

Parity

no parity

no parity

No

 

The desired baud rate can be configured using the „K114Config“ software.

Further connection parameters are possible for communicating via the USB interface with the RS485.

Property

USB / RS485

configurable

Baud rate

freely selectable, up to 250 kbps

Yes

Data bits

7,8

Yes

Stop bits

1,2

Yes

Parity

Odd, even, mark, space, no parity

Yes

 

The K-114 BT converter communicates with the connected devices via an RS485 (half-duplex mode) bus. The 

internal battery supplies power to the devices connected to the K-114 BT. KELLER products operate with fail-safe 

drivers, which output a logical high at the receiver outlet in the event of short-circuited, open or non-terminated 

 inputs in order to avoid invalid signal states. KELLER products also feature slew rate limitation, which limits the edge 

steepness of the driver output. This prevents high-frequency emissions from devices and data lines. A maximum of 

128 bus devices can be connected to this RS485 master.

Summary of Contents for K-114 BT

Page 1: ...nterface converter KELLER AG St Gallerstrasse 119 CH 8404 Winterthur Tel 41 0 52 235 25 25 Fax 41 0 52 235 25 00 KELLER GmbH Schwarzwaldstrasse 17 D 79798 Jestetten Tel 49 0 7745 9214 0 Fax 49 0 7745...

Page 2: ...ster PC must support Bluetooth Classic 2 1 or higher Function When the device is switched on the Bluetooth module is activated automatically and the interface converter is visi ble to other Bluetooth...

Page 3: ...deactivated it is again possible to communicate via the USB interface Pairing connecting via Bluetooth for the first time The computer must have an internal Bluetooth module or a Bluetooth USB dongle...

Page 4: ...x W x D 95 x 28 x 65 mm Weight 150 g Battery charging In order to protect the battery in the device the user should switch off the converter after use In battery operation the K 114 BT switches off a...

Page 5: ...ansmission speed must be configu red on the master PC and on the K 114 BT Property Bluetooth RS485 USB internal controller bus address 253 configurable Baud rate 9600 default 115200 baud 9600 default...

Page 6: ...ed Termination resistor not activated Caution In battery operated end devices this function can lead to errors due to the increased power load Recommendation do not activate this function 300 ms LED d...

Page 7: ...port red off gr n green gr n flashing POWER LED Check install K 114 BT software driver Internal error Device is out of operation it may have switched itself off Press the Power Button to turn the devi...

Page 8: ...gnostic tool for the converter K 114 BT It can be used to configure the baud rate of the converter as well as to select the echo function and to switch on the bias resistors or the termination resisto...

Page 9: ...acity of the battery drops below 20 the user is notified that the battery is low on power Battery low connect charger This final warning appears when the capacity drops below 10 and instructs the user...

Page 10: ...chtlich elektromagnetischer Vertr g lichkeit wurde folgende Normen heran gezogen Jestetten 26 Mai May mai 2016 Hannes W Keller Gesch ftsf hrender Inhaber Managing Owner Pr sident Directeur G n ral mit...