KELLER
May 2006
Page 1: ...High Pressure Calibrator Standard and Full Version KELLER May 2006...
Page 2: ...ble RS 232 2 Program Discs Logger Windows Logger MS DOS Included in Delivery High Pressure Calibrator Full Version 1 Calibrator 1 Carrying Case 1 Test Unit Adapter G 1 4 G 1 2 1 Test Unit Adapter G 1...
Page 3: ...ing of Pressure Switches 9 Pst Mode Preparing Storage of Measured Values 11 Rec Mode Executing Storage of Measured Values 12 Rec Mode Measuring the Linearity 13 Lin Mode Calculating the Linearity 14 L...
Page 4: ...Refill Screw 10 Oil Screw Compressor 11a Outlet Pressure Port G 1 4 female with Hexagon Plug 200 bar and 350 bar Calibrator with Overpressure Valve 11b Outlet Pressure Port G 1 4 female with Hexagon P...
Page 5: ...essure with both the integrated priming pump and the oil screw compressor The pressure can be fine adjusted The high technology integrated in this instru ment allows it to measure precisely both the p...
Page 6: ...within the oil chamber the oil refill screw 9 must be open when operating the calibrator Generate pressure with the priming pump 12 up to approx 10 bar Increase or reduce pressure with screw com press...
Page 7: ...Mano Release the F Key Within specific functions the F Key can execute additional commands Exe Step Reset Activating Off turns off the instrument When starting the instrument again the display will au...
Page 8: ...ay shows the tem perature Pressing the Hold Key H Key freezes the measured values The actual pressure is now shown on the lower right display Pressing the H Key again leads back into the Mano Mode wit...
Page 9: ...e With the command Exe the new zero is being determined and stays as a reference until a new Tara is being executed or until the instrument is turned off The program returns to the Mano Mode The Unit...
Page 10: ...sure Activate Sel Flashing Exe changes to hr hours Exe again to day Activate Step First digit is flashing Exe encreases digit With Step to next digit Activate Step Second digit is flashing Exe encreas...
Page 11: ...o return to the Prep Mode from the End display A new test can be started as follows also when measurements are running Press the H Key or Activate Run with the F Key At any time you may return to the...
Page 12: ...ing recorded and can be written into a PC and displayed on the screen only with Logger Option Activating End followed by the activation of another function quits the Pst Mode and leads into the new se...
Page 13: ...Low if the measurement started at high pressure and the pressure switch was switching at a lower pressure An increase of pressure from this status writes the new switch point automatically into the sm...
Page 14: ...er recording the storage requirements are tripling 1000 measurements with three values each lead to 3000 measuring values Adjustment of the Storage Parameters Activate Rec Recording interval Number of...
Page 15: ...ay At any time a record can prematurely be termi nated by activating Res In the Rec Mode you may also display the tem perature Pressing the H key at Prep display indicates the temperature on ther lowe...
Page 16: ...Maximal 10 linearity values are allowed After having activated Lin activating the com mand Save writes the actual values of the reference pressure and of the test unit into the memory This cycle can...
Page 17: ...ersion only From this position the linearity of all other measuring points can be called upon Pressure at point 3 Activate Exe Pressure at point 4 14 Number of last measuring point Activate Calc Actua...
Page 18: ...endency of posi tion or the influence of temperature can result in zero shifts These shifts can be permanently corrected as follows Activate Zero Actual Zero Activate Exe Activate Reso Activate Reso T...
Page 19: ...e absolute time axis is calculated from the time set in the PC the time meter of the calibrator and the recorded time The calibrator should therefore not be turned off between the recor ding and the t...
Page 20: ...necting your test unit Open the battery compartment of the instrument base and change the batteries Restarting the Calibrator If it should occur that the programme is locked meaning that the instrumen...
Page 21: ...transmitter in mA or V Connecting Pressure Switches Plug in the pressure switch adapter Internal supply is OK when LED 3 lights Connect the pressure switch to adapter FUNCTIONS Mano Display of pressu...
Page 22: ...otal Oil Chamber Volume Oil Type Dead Volume Total Weight including Batteries Dimensions L x W x H Protection only Standard 0 1 FS Full Scale from 0 50 C 0 1 FS 1 Digit 10 mbar 100 mbar 100 mbar FS 10...
Page 23: ...ion of which a drawing marked with the CE sign exists and which are produced according to this drawing This explanation has no validity without such a drawing As criteria for the electromagnetic compa...