SCPI Command Reference
Models 2510 and 2510-AT User’s Manual
Read version of SCPI standard
:VERSion?
:SYSTem:VERSion?
Read SCPI version
Description
This query command is used to read the version of the SCPI standard
being used by the Model 2510. Example code:
1996.0
The above response message indicates the version of the SCPI standard.
Reset timestamp
:RESet
:SYSTem:TIME:RESet Reset
timestamp
Description
This action command is used to reset the absolute timestamp to 0 sec-
onds. The timestamp also resets to 0 seconds every time the Model 2510
is turned on. Note that time information can be returned with the
:READ? and :MEAS? queries by specifying a time element with the
:FORMat:ELEMents command (see
2-wire/4-wire sense mode
:RSENse <b>
:SYSTem:RSENse <b>
Control 4-wire/2-wire sensing
Parameters
<b> =
1 or ON
Enable 4-wire sensor sense mode
0 or OFF
Disable 4-wire sensing (2-wire mode)
Query
:RSENse?
Query enabled/disabled 4-wire sense state
Description
:RSENs enables or disables 4-wire sensing for RTD and thermistor tem-
perature transducers. 4-wire sensing should be used for RTD and ther-
mistor sensors with <1k
Ω
resistance.