Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-308
2600BS-901-01 Rev. B / May 2013
Details
These attributes are used to read or write to the questionable status instrument summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register bits are set.
The least significant bit of the binary number is bit B0, and the most significant bit is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status register set contents
(on
page E-1) and
Enable and transition registers
(on page E-19). The individual bits of this register are defined in
the following table.
Bit
Value
Description
B0
Not used
Not applicable.
B1
status.questionable.instrument.SMUA
Set bit indicates one or more enabled bits for the
SMU A questionable register are set.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.questionable.instrument.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set bit
indicates one or more enabled bits for the SMU B
questionable register are set.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15
Not used
Not applicable.
As an example, to set bit B1 of the questionable status instrument summary enable register, set
status.questionable.instrument.enable = status.questionable.instrument.SMUA
.
In addition to the above constants,
questionableRegister
can be set to the numeric equivalent of the bit to
set. To set more than one bit of the register, set
operationRegister
to the sum of their decimal weights. For
example, to set bits B1 and B2, set
questionableRegister
to 6 (which is the sum of 2 + 4).
Example
status.questionable.instrument.enable =
status.questionable.instrument.SMUA
Uses a constant to set the SMUA bit
of the questionable status
instrument summary enable register.
Also see
Questionable Status Registers
(on page E-13)
status.questionable.*
(on page 7-304)
status.questionable.instrument.smuX.*
This attribute contains the questionable status SMU X summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
4864 (All bits set)