Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-258
2600BS-901-01 Rev. B / May 2013
status.measurement.buffer_available.*
This attribute contains the measurement event buffer available summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
Models 2601B/2611B/2635B: 2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B:
6 (All bits set)
Usage
measurementRegister
= status.measurement.buffer_available.condition
measurementRegister
= status.measurement.buffer_available.enable
measurementRegister
= status.measurement.buffer_available.event
measurementRegister
= status.measurement.buffer_available.ntr
measurementRegister
= status.measurement.buffer_available.ptr
status.measurement.buffer_available.enable =
measurementRegister
status.measurement.buffer_available.ntr =
measurementRegister
status.measurement.buffer_available.ptr =
measurementRegister
measurementRegister
The status of the measurement event register; a zero (0) indicates no bits set
(also send 0 to clear all bits); other values indicate various bit settings
Details
These attributes are used to read or write to the measurement event buffer available summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which register
bits are set. The least significant bit of the binary number is bit B0, and the most significant bit is bit B15. For
example, assume value 6 is returned for the enable register. The binary equivalent is 0000 0000 0000 0110. This
value indicates that bit B1 (SMUA) and bit B2 (SMUB) are set.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status register set contents
(on
page E-1) and
Enable and transition registers
(on page E-19). The individual bits of this register are defined in
the following table.
Bit
Value
Description
B0
Not used
Not applicable.
B1
status.measurement.buffer_available.SMUA
Set bit indicates that there is at least one
reading stored in either or both of the dedicated
reading buffers.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.measurement.buffer_available.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set
bit indicates that there is at least one reading
stored in either or both of the dedicated reading
buffers.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15
Not used
Not applicable.
Summary of Contents for System SourceMeter 2601B