Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-228
2600BS-901-01 Rev. B / May 2013
smuX.source.output
This attribute sets the source output state (on or off).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0 (smuX.OUTPUT_OFF)
Usage
sourceOutput
= smu
X
.source.output
smu
X
.source.output =
sourceOutput
sourceOutput
The source's output state setting; set to one of the following values:
0
or
smu
X
.OUTPUT_OFF
: Turns off the source output
1
or
smu
X
.OUTPUT_ON
: Turns on the source output
2
or
smu
X
.OUTPUT_HIGH_Z
: Turns off the output in high Z mode (allows you to go
to high Z mode without first setting the
smu
X
.source.offmode
attribute to
smu
X
.OUTPUT_HIGH_Z
)
X
Source-measure unit (SMU) channel (for example,
smua.source.output
applies
to SMU channel A)
Details
Reading this attribute returns the output state of the source. Setting this attribute switches the output of the
source on or off.
When the output is switched on, the SMU sources either voltage or current, as set by
smu
X
.source.func
.
Setting this attribute to
smu
X
.OUTPUT_HIGH_Z
causes the output to turn off and go to the High Z mode. If the
smu
X
.source.output
is read after setting this attribute to
smu
X
.OUTPUT_HIGH_Z
, it returns
0
.
Example
smua.source.output = smua.OUTPUT_ON
Turns on SMU channel A source
output.
Also see
DUT test connections
(on page 2-48)
smuX.source.func
(on page 7-221)
smuX.source.offmode
(on page 7-226)
smuX.source.outputenableaction
This attribute controls output enable action of the source.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
0 (smuX.OE_NONE)