Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-223
Details
This attribute configures the output level of the voltage or current source.
If the source is configured as a voltage source and the output is on, the new
smu
X
.source.levelv
setting is
sourced immediately. If the output is off or the source is configured as a current source, the voltage level is
sourced when the source is configured as a voltage source and the output is turned on.
If the source is configured as a current source and the output is on, the new
smu
X
.source.leveli
setting is
sourced immediately. If the output is off or the source is configured as a voltage source, the current level is
sourced when the source is configured as a current source and the output is turned on.
The sign of
sourceLevel
dictates the polarity of the source. Positive values generate positive voltage or current
from the high terminal of the source relative to the low terminal. Negative values generate negative voltage or
current from the high terminal of the source relative to the low terminal.
The
reset()
function sets the source levels to 0 V and 0 A.
Example
smua.source.levelv = 1
Sets voltage source of SMU
channel A to 1 V.
Also see
smuX.source.compliance
smuX.source.func
(on page 7-221)
smuX.source.output
(on page 7-228)
Source-measure concepts
(on page 4-1)
smuX.source.limitY
This attribute sets compliance limits.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
Limit voltage:
Models 2601B/2602B/2604B: 40 (40 V)
Models
2611B/2612B/2614B/2634B/2635B/2636B:
20 (20 V)
Limit current:
Models 2601B/2602B/2604B: 1 (1 A)
Models
2611B/2612B/2614B/2634B/2635B/2636B:
100e-3 (100 mA)
Limit power: 0 (disabled)
Usage
limit
= smu
X
.source.limit
Y
smu
X
.source.limit
Y
=
limit
limit
The compliance limit value; set to one of the following values:
Voltage compliance:
Models 2601B/2602B/2604B: 10 mV to 40 V
Models 2611B/2612B/2614B/2634B/2635B/2636B: 20 mV to 200 V
Current compliance:
Models 2601B/2602B/2604B/2611B/2612B/2614B: 10 nA to 3 A
Models 2634B/2635B/2636B: 100 pA to 1.5 A
Power compliance (in watts)
X
Source-measure unit (SMU) channel (for example,
smua.source.limitv
applies to SMU
channel A)
Y
SMU function (
v
= voltage,
i
= current)
Summary of Contents for System SourceMeter 2601B