The main display screen reappears:
3. Press the CURSOR
keys to move the cursor to the
10s digit.
4. Press the navigation wheel
to enter EDIT mode. The EDIT
indicator appears in the upper left corner of the display.
5. Turn the navigation wheel
to set the source value to
20.0000 V
, and then press the navigation wheel
to enter the
selection and exit EDIT mode.
Step 2: Set the source limit
1. Press the
LIMIT
key. You will see a blinking character in the
LimA value field.
2. While that character is still blinking, press the down
RANGE
keys as needed to select the
10 mA
limit range.
Verify that the source limit value in the LimA field is
10.0000 mA.