12-72
Return to
2600S-901-01 Rev. C / January 2008
Section 12: Instrument Control Library
Series 2600 System SourceMeter® Instruments Reference Manual
smuX.source.levelY
X = SMU channel (a or b)
Y = SMU measure function (v or i)
where v = voltage, i = current
Attribute
Source levels.
Usage
sourceval = smuX.source.levelY
-- Reads source value.
smuX.source.levelY = sourceval
-- Writes source value.
sourceval
Set 2601/2602 voltage from 0 to ±40 (volts).
Set 2601/2602 current from 0 to ±3 (amps).
Set 2611/2612/2635/2636 voltage from 0 to
±200 (volts).
Set 2611/2612/2635/2636 current from 0 to
±1.5 (amps).
Remarks
• This attribute configures the source level of the voltage or current source.
• If the source is configured as a voltage source and the output is on, the new
smuX.source.levelv
setting will be sourced immediately. If the output is off or if the
source is configured as a current source, the voltage level will be sourced when the
source is configured as a voltage source and the output is turned on.
• If the source is configured as a current source and the output is on, the new
smuX.source.leveli
setting will be sourced immediately. If the output is off or if the
source is configured as a voltage source, the current level will be sourced when the
source is configured as a current source and the output is turned on.
• The sign of
sourceval
dictates the polarity of the source.
Positive values generate
positive voltage or current from the high terminal of the source relative to the low
terminal. Negative values generate negative voltage or current from the high
terminal of the source relative to the low terminal.
• The
reset
function sets the source levels to 0V and 0A.
Details
See
"basic operation"
Also see
,
Example
Sets V-source to 1V for SMU A:
smua.source.levelv = 1
smuX.source.limitY
X = SMU channel (a or b)
Y = SMU measure function (v or i)
where v = voltage, i = current
Attribute
Compliance limits.
Usage
limit = smuX.source.limitY
-- Reads compliance limit.
smuX.source.limitY = limit
-- Writes compliance limit.
limit
2601/2602 voltage compliance from 0 to ±40 (volts).
2601/2602 current compliance from 0 to ±3 (amps).
2611/2612/2635/2636 voltage compliance from 0 to
±200 (volts).
2611/2612/2635/2636 current compliance from 0 to
±1.5 (amps).
Remarks
• Use the
smuX.source.limiti
attribute to limit the current output of the voltage
source. Use
smuX.source.limitv
to limit the voltage output of the current source.
The SMU will always choose (auto range) the source range for the limit setting.
• This attribute should be set in the test sequence before the turning the source on.
• Reading this attribute indicates the presently set compliance value. Use
smuX.source.compliance
to read the state of source compliance.
Details
See
"basic operation".
Also see
,
Example
Sets V-compliance to 30V for SMU A:
smua.source.limitv = 30