Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-154
2600BS-901-01 Rev. C / August 2016
Example
localnode.password = "N3wpa55w0rd"
Changes the remote interface password to
N3wpa55w0rd.
Also see
(on page 7-132)
(on page 7-154)
localnode.passwordmode
This attribute stores the remote access password enable mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
1 (localnode.PASSWORD_WEB)
Usage
mode
= localnode.passwordmode
localnode.passwordmode =
mode
mode
The remote password enable mode
Details
This attribute controls if and where remote access passwords are required. Set this attribute to one of
the values below to enable password checking:
localnode.PASSWORD_NONE
or
0
: Disable passwords everywhere
localnode.PASSWORD_WEB
or
1
: Use passwords on the web interface only
localnode.PASSWORD_LAN
or
2
: Use passwords on the web interface and all LAN interfaces
localnode.PASSWORD_ALL
or
3
: Use passwords on the web interface and all remote command
interfaces
When using this command from a remote node,
localnode
should be replaced with the node
reference, for example
node[5].passwordmode
.
Example
mode = localnode.PASSWORD_WEB
localnode.passwordmode = mode
Sets value of
mode
to
PASSWORD_WEB
.
Allows use of passwords on the web interface only.
Also see
localnode.prompts
This attribute sets and reads the local node prompting state (enabled or disabled).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Power cycle
Not saved
0 (disabled)