Series 2600 System SourceMeters Reference Manual
Status Model
D-9
Return to
2600S-901-01 Rev. A / May 2006
Clearing registers and queues
When the SourceMeter is turned on, various register status elements will be set
as follows:
• The PON bit in the
status.condition
register will be set.
• Bits such as the output enable and over-temperature bits will be set
appropriately.
• All enable registers will be set to 0.
• All NTR registers will be set to 0.
• All used PTR register bits will be set to 1.
• The two queues will be empty.
Commands to reset the status registers and the Error Queue are listed in
. In addition to these commands, any programmable register can be reset by send-
ing the 0 parameter value with the individual command to program the register.
Table D-2
Commands to reset registers and clear queues
Commands
Description
To Reset Registers:
*CLS
or
status.reset()
Reset bits of status registers to 0.
To Clear Error Queue:
errorqueue.clear()
Clear all messages from Error Queue.
System events
status.system.enable
status.system2.enable
status.system3.enable
status.system4.enable
status.system5.enable
Table D-1 (cont.)
Status functions and registers
Type
Function
1