Table 15-4
Model 2601/2602 voltage measurement accuracy limits
Model 2601/2602 source
and measure range
1
Source voltage
2
Model 2601/2602 voltage
reading limits
(1 year, 18°C–28°C)
100mV
90.000mV
89.8365 to 90.1635mV
1V
0.90000V
0.899665 to 0.900335V
6V
5.4000V
5.39819 to 5.40181V
40V
36.000V
35.9866 to 36.0134V
1. Measure range coupled to source range when simultaneously sourcing and measuring voltage.
2. As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Table 15-5
Model 2611/2612 voltage measurement accuracy limits
Model 2611/2612 source
and measure range
1
Source voltage
2
Model 2611/2612 voltage
reading limits
(1 year, 18°C–28°C)
200mV
180.000mV
179.748 to 180.252mV
2V
1.80000V
1.79938 to 1.80062V
20V
18.0000V
17.9923 to 18,0077V
200V
180.000V
179.923 to 180.077V
1. Measure range coupled to source range when simultaneously sourcing and measuring voltage.
2. As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Series 2600 System SourceMeters Reference Manual
Performance Verification 15-11
Return to
2600S-901-01 Rev. A / May 2006