Series 2600 System SourceMeters Reference Manual
Factory Scripts 13-11
Return to
2600S-901-01 Rev. A / May 2006
TSP Project Name:
KIFactoryGeneral
TSP Test Script Name:
KIGeneral
Firmware Version:
1.0.2 and later
Function:
SweepIListMeasureV(smu, ilist, stime, points)
Description: Performs a current list sweep with voltage measured at every step (point):
• Sets the
smu
to output the first
ilist
amps value, allows the source to
settle for
stime
seconds and then performs a voltage measurement.
• Sets the
smu
to output the next
ilist
amps value, allows the source to
settle for
stime
seconds and then performs a voltage measurement.
• Repeats the above sequence until the voltage is measured for the last
amps value. The last point in the list to be measured is
points
.
Parameters:
smu, ilist, stime, points
smu:
SourceMeter Channel (A or B). Defaults to smua if all parameters
are omitted when function is called.
ilist:
Arbitrary list of current source values:
ilist = {value1, value2, ....valueN}
stime:
Settling time in seconds. Occurs after sourcing a value and before
performing a measurement.
points:
Number of sweep measure points (must be
≥
2).
Data: Voltage measurements, current source values and timestamps are stored in
smuX.nvbuffer1
.
Example:
myilist = {-100E-9, 100E-9, -1E-6, 1E-6, -1E-3, 1E-3}
SweepIListMeasureV(smua, myilist, 500E-6, 6)
This function performs a 6-point current list sweep starting at the first point in
myilist
. Voltage is measured at every step (point) in the sweep. The source
will be allowed to settle on each value for 500µs before a measurement is
performed.