3-12 DUT Test Connections
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Figure 3-11
HI
LO
HI
LO
DUT
HI
LO
Metal Safety
Shield
SMU A
2601/2602-1
SMU A
2601/2602-2
Safety
Earth
Ground
40V
40V
80V
Safety shield for hazardous voltage using two 2601/2602 channels (>42V)
Figure 3-12
CHANNEL A
LO
G
LO
HI G G G HI
S
S
Model 2601/2602-1
DUT
HI
LO
CHANNEL A
LO
G
LO
HI G G G HI
S
S
Model 2601/2602-2
LO
HI
Safety
Earth
Ground
Chassis
Chassis
Screw
Connections for test circuit shown in
Model 2611/2612 safety shield
The maximum output voltage for a Model 2611/2612 channel is 200V, which is
considered hazardous and requires a safety shield (
). The connections
for the test configuration in
are shown in
. Use #18AWG
wire or larger for connections to safety earth ground and chassis.