4-8
Troubleshooting
plexer directs the voltage feedback, current feedback, reference, or ground signal to the
A/D converter. An opto-isolated interface provides control signals for both DACs, analog
circuit control, and A/D converter communication to the digital section.
Power supply
Figure 4-3 shows a block diagram of the Model 2425 power delivery system.
The offline switching power supply provides all power for the instrument while providing
universal inputs and power factor correction for the 120/240V line. The digital board runs
off of 5V and 12V supplies derived from the switcher. The +12VD supply is set to pro-
gram the flash ROM. (See Digital circuitry below.)
A constant-frequency switching supply runs off the +24VD supplies and generates all the
floating and output supply voltages for the analog board: +5V and ±15V, ±42V and
±150V.
Figure 4-3
Power supply block diagram
Analog Board
+15V
+5V
F
-15V
F
F
F
Output Stage
-150V -42V
+42V +150V
O
Constant Frequency
Low Noise Floating
Switching Supply
+24Vdc
D
Line
Neutral
+5Vdc
D
+12Vdc
PFC
DC/DC
Converter
D
Digital
Circuits
+24dc
DC
Regulators
Summary of Contents for SourceMeter 2425
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 35: ...1 20 Performance Verification Model 2425 Service Manual...
Page 36: ...2 Calibration...
Page 62: ...3 RoutineMaintenance...
Page 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Page 66: ...4 Troubleshooting...
Page 84: ...5 Disassembly...
Page 92: ...6 ReplacementParts...
Page 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Page 114: ...A Specifications...
Page 116: ...B CommandReference...
Page 128: ...C CalibrationPrograms...
Page 139: ......
Page 141: ......