Series 2200 Programmable DC Power Supplies Reference Manual
Section 4: Command reference
2200S-901-01 Rev. C / March 2016
4-13
[SOURce:]FUNCtion:MODE
This command can be in either fixed mode or list mode. When this
command is in fixed mode, the power supply responds to discrete
commands. When this command is in list mode, the power supply
operates in list mode.
Fixed mode
Set this command to
FIXed
to edit a list file. Once the list is
finalized and you are ready to use the list, set this command to
LIST
to operate in list mode.
List mode
Set this command to
LIST
to operate the power supply in list
mode. Attempts to edit a list while this command is in list mode will
generate an error message of
settings conflict
' If you need
to change the list, set this command to
FIXed
before changing the
list.
Group
Source
Syntax
[SOURce:]FUNCtion:MODE {FIXed|LIST}
[SOURce:]FUNCtion:MODE?
Arguments
FIXed
configures the power supply to respond to discrete
commands.
LIST
configures the power supply to operate in list mode.
Returns
FIX|LIST
Examples
[SOURCE:]FUNCTION:MODE LIST
[SOURce:]LIST:COUNt
This command configures the number of times the active list will
execute before stopping.
Group
Source
Syntax
[SOURce:]LIST:COUNt {<NR1>|ONCE|REPeat}
[SOURce:]LIST:COUNt?
Related
Commands
(on page 4-15)
(on page 4-13)
Arguments
<NR1>
is an integer between
2
and
65535
. It determines the
number of times the active list will execute.
ONCe: configures the lists to execute once and stop.
REPeat: configures the lists to loop continuously.