Principles of Operation
17
2
2
2
2
2
2
2
2
2
Specifying Analog Output Channels
You can output data continuously from one or both analog output
channels using the output channel list. If you want to output data
from one analog output channel continuously, specify either 0
(DAC0) or 1 (DAC1) in the channel output list. If you want to output
data to both analog output channels continuously, specify the output
channel list in the following order: 0, 1.
Then, use software to specify the data flow mode as continuous for
the D/A subsystem; refer to
for more information on
continuous analog output operations.
Output Resolution
The KUSB-3100 module provides a fixed output resolution of 12 bits.
Note that the resolution is fixed; it cannot be programmed in
software.
Output Ranges and Gains
The KUSB-3100 module provides a fixed output range of ±10 V (the
gain is ignored).
Output Trigger
A trigger is an event that occurs based on a specified set of
conditions. The KUSB-3100 module supports a software trigger for
starting analog output operations. The module starts outputting data
when it receives a software command (trigger).
Use software to specify the trigger source for the D/A subsystem as a
software trigger.
Summary of Contents for KUSB-3100
Page 10: ...Contents x...
Page 14: ...About this Manual xiv...
Page 15: ...1 1 Overview Features 2 Supported Software 3...
Page 18: ...Chapter 1 4...
Page 46: ...Chapter 2 32...
Page 47: ...33 3 Supported Device Driver Capabilities...
Page 86: ...Chapter 4 72...
Page 87: ...73 5 Troubleshooting General Checklist 74 Service and Support 77...
Page 100: ...Chapter 6 86...
Page 101: ...87 A Specifications...
Page 112: ...Appendix A 98...
Page 113: ...99 B Connector Pin Assignments...
Page 115: ...101 C Reading from or Writing to the Digital Registers...