Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-153
smu.source.userdelay[N]
This attribute sets a user-defined delay that can be used in the trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
0
Usage
delayTime
= smu.source.userdelay[
N
]
smu.source.userdelay[
N
] =
delayTime
delayTime
The delay in seconds (0 to 10000)
N
The user delay to which this time applies (1 to 5)
Details
To use this commands in a trigger model, assign the delay to the dynamic delay block.
Example
smu.source.userdelay[1] = 3
Set user delay 1 for the source to be
3 seconds.
Also see
trigger.model.setblock() — trigger.BLOCK_DELAY_DYNAMIC
(on page 8-206)
status.clear()
This function clears event registers and the event log.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
status.clear()
Details
This command clears the event registers of the Questionable Event and Operation Event Register set. It does
not affect the Questionable Event Enable or Operation Event Enable registers.
Example
status.clear()
Clear the bits in the registers
Also see
(on page B-2)