Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-344
DMM7510-901-01 Rev. B / May 2015
trigger.tsplinkout[N].stimulus
This attribute specifies the event that causes the synchronization line to assert a trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
TSP-Link line
N
reset
Configuration script
trigger.EVENT_NONE
Usage
event
= trigger.tsplinkout[
N
].stimulus
trigger.tsplinkout[
N
].stimulus =
event
event
The event identifier for the triggering event (see
Details
)
N
The trigger line (1 to 3)
Details
To disable automatic trigger assertion on the synchronization line, set this attribute to
trigger.EVENT_NONE
.
Do not use this attribute when triggering under script control. Use
trigger.tsplinkout[
N
].assert()
instead.
The
event
parameters that you can use are described in the table below.
Trigger events
Event description
Event constant
No trigger event
trigger.EVENT_NONE
Front-panel TRIGGER key press
trigger.EVENT_DISPLAY
Notify trigger block
N
(1 to 8) generates a trigger event when the
trigger model executes it
trigger.EVENT_NOTIFY
N
A command interface trigger (bus trigger):
•
Any remote interface: *
TRG
•
GPIB only: GET bus command
•
VXI-11: VXI-11 command
device_trigger
trigger.EVENT_COMMAND
Line edge (either rising, falling, or either based on the configuration of
the line) detected on digital input line
N
(1 to 6)
trigger.EVENT_DIGIO
N
Line edge detected on TSP-Link synchronization line
N
(1 to 3)
trigger.EVENT_TSPLINK
N
Appropriate LXI trigger packet is received on LAN trigger object
N
(1
to 8)
trigger.EVENT_LAN
N