Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-286
DMM7510-901-01 Rev. B / May 2015
trigger.model.load() — Config List
This function loads a predefined trigger model configuration that uses a measure configuration list.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.model.load("ConfigList",
measureConfigList
)
trigger.model.load("ConfigList",
measureConfigList, delay
)
trigger.model.load("ConfigList",
measureConfigList
,
delay
,
bufferName
)
trigger.model.load("ConfigList",
measureConfigList,
delay
,
bufferName,
readingBlock
)
measureConfigList
A string that contains the name of the measurement configuration list to
use
delay
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
bufferName
The name of the reading buffer, which may be a default buffer
(
defbuffer1
or
defbuffer2
) or a user-defined buffer; defaults to
defbuffer1
.
readingBlock
Define a measure or digitize block for the trigger model; options are:
•
trigger.READING_ACTIVE
: Add a measure or digitize block to the
trigger model based on the active function; if no option defined,
trigger.READING_ACTIVE
is used
•
trigger.READING_MEASURE
: Adds a measure block to the trigger
model
•
trigger.READING_DIGITIZE
: Adds a digitize block to the trigger
model
Details
This trigger model template incorporates a configuration list. You must set up the configuration lists
before loading the trigger model.
You can also set a delay and change the reading buffer.
The out line of the EXT TRIG IN/OUT rear-panel terminal is asserted at the end of each
measurement.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel
MENU
key and under Trigger, selecting
Configure
. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist()
command to view the trigger model blocks in a list format.