Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-12
DMM7510-901-01 Rev. B / May 2015
acal.revert()
This function returns auto calibration constants to the previous constants.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
acal.revert()
Details
This command reverts the present set of auto calibration constants to the previous set of auto
calibration constants.
The last run time and internal temperature are reverted to the previous values. The auto calibration
count is not changed.
Example
acal.revert()
Auto calibration values are reverted to the
previous set of auto calibration constants.
Also see
(on page 8-12)
acal.run()
This function immediately runs auto calibration and stores the constants.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
acal.run()
Details
During auto calibration, a progress message is displayed on the front panel. At completion, an event
message is generated.
If you have set up auto calibration to run at a scheduled interval, when you send the run command,
the instrument adjusts the next scheduled auto calibration to be the next interval. For example, if auto
calibration is scheduled to run every 7 days, but you run auto calibration on day 3, the next auto
calibration will run 7 days after day 3.
Example
acal.run()
Auto calibration starts running. When it is complete, an information message is
generated in the event log.
Also see
(on page 8-13)
(on page 3-44)
(on page 8-221)