Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-173
:TRACe:WRITe:FORMat
This command sets the units and number of digits of the readings that are written into the reading buffer.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRACe:WRITe:FORMat "<bufferName>", <units>, <displayDigits>
:TRACe:WRITe:FORMat "<bufferName>", <units>, <displayDigits>, <extraUnits>
:TRACe:WRITe:FORMat "<bufferName>", <units>, <displayDigits>, <extraUnits>,
<extraDigits>
<bufferName>
A user-supplied string that indicates the name of the buffer
<units>
The units for the first measurement in
the buffer index:
•
AMP
•
AMP_AC
•
CELSius
•
DECibel
•
FAHRenheit
•
FARad
•
HERTz
•
KELVin
•
NONE
•
OHM
•
PERCent
•
RATio
•
RECiprocal
•
SECond
•
VOLT
•
VOLT_AC
•
WATT
•
X
<displayDigits>
Integer from 3 to 8
<extraUnits>
The units for the second measurement in the buffer index; the selections are
the same as
<units>
(only valid for buffer style
FULLWRITable
); if this
parameter is not specified, the value for
<units>
is used
<extraDigits>
The number of digits to use for the second measurement; the selections are the
same as
<displayDigits>
(only valid for buffer style
FULLWRITable
); if this
parameter is not specified, the value for
<displayDigits>
is used
Details
This command is valid when the buffer style is writable or full writable.
Defines the units and the number of digits that are reported for the data. This function affects how the
data is shown in the reading buffer and what is shown on the front-panel Home, Histogram, Reading
Table, and Graph screens.