Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-20
DMM7510-901-01 Rev. B / May 2015
:ACAL:RUN
This command immediately runs auto calibration and stores the constants.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
Not applicable
Usage
:ACAL:RUN
Details
During auto calibration, a progress message is displayed on the front panel. At completion, an event
message is generated.
If you have set up auto calibration to run at a scheduled interval, when you send the run command,
the instrument adjusts the next scheduled auto calibration to be the next interval. For example, if auto
calibration is scheduled to run every 7 days, but you run auto calibration on day 3, the next auto
calibration will run 7 days after day 3.
Example
ACAL:RUN
Auto calibration starts running.
Also see
(on page 3-44)