Model 6517 Connections
Refer to Figure 2-3 to connect the Model 6517 to the Model 8009 test fixture.
The triax cable and the Model 6517-ILC-3 interlock cable are supplied with the
Model 8009. Note that the ground link on the Model 6517 must be removed.
Proper grounding will be performed by the Model 8009.
2.5
Model 6517B Connections
Refer to Figure 2-4 to connect the Model 6517B to the Model 8009 test fixture.
The triax cable and the 6517B-ILC-3 interlock cable are supplied with the Model
8009. Note that the ground link on the Model 6517B must be removed. Proper
grounding will be performed by the Model 8009.
Figure 2-3
Connecting the Model 6517 Electrometer/High Resistivity Meter to the Model
8009 test fixture