41
CALCulate3:LIMit[1]:DIGital
Path to set digital limit 1attributes
CALCulate3:LIMit2:DIGital
Path to set digital limit 2attributes
:MODE <name> [, <clist>]
Select digital limit mode
:MODE? [, <clist>]
Query digital limit mode
For non-scan channels, the :MODE command is global to all digital input channels. That is, the selected mode (PATTern or
MASK) applies to all digital input channels.
When configuring a scan, use the <clist> parameter. Each digital input channel can have its own unique digital limits
configuration. For example, digital input channel 111 can be set for the pattern mode, while input channel 112 can be set for
the mask mode.
With the MODE? command, you can read the selected global mode or the mode of individual input scan channels using the
<clist> parameter.
Errors: -221 settings conflict – Tried to set the mode for a digital output scan channel.
-224 Illegal parameter value – Tried to set the mode for an analog input scan channel.
Examples:
CALC3:LIM1:DIG:MODE MASK,(@111)
‘ Set limit 1 mask mode for scan channel.
CALC3:LIM1:DIG:MODE PATT
‘ Set limit 1 global pattern mode.
CALC3:LIM1:DIG:MODE?,(@111)
‘ Query limit 1 mode for scan channel.
CALC3:LIM1:DIG:MODE?
‘ Query limit 1 global mode.
:PATTern <NDN or NRf> [, <clist>]
Set digital limit pattern
:PATTern? [, <clist>]
Query digital limit pattern
For non-scan channels, the :PATTern command is global to all digital input channels. That is, the set bit pattern applies to all
digital input channels.
When configuring a scan, use the <clist> parameter. Each digital input channel can have its own unique digital limits
configuration. For example, a pattern can be set for digital input channel 111, while a different pattern can be set for input
channel 112.
With the PATTern? command, you can read the set global pattern or the pattern of individual input scan channels using the
<clist> parameter. The format for the returned value(s) are determined by the selected digital data format (see :FORMat
command).
Errors: -221 settings conflict – Tried to set a pattern for a digital output scan channel.
-224 Illegal parameter value – Tried to set a pattern for an analog input scan channel.
Examples:
CALC3:LIM1:DIG:PATT #B100101,(@111)
‘ Set pattern for scan channel; binary format.
CALC3:LIM1:DIG:PATT 37
‘ Set global pattern; decimal format.
CALC3:LIM1:DIG:PATT #H25,(@111)
‘ Set pattern for scan channel; hexadecimal format.
CALC3:LIM1:DIG:PATT #H25
‘ Set global pattern; hexadecimal format.
CALC3:LIM1:DIG:PATT?
‘ Query global pattern.
CALC3:LIM1:DIG:PATT? (@111)
‘ Query pattern for scan channel.