45
:ARM[:SEQuence[1]]
:LAYer2
:SOURce?
:TIMer <n>
:TIMer?
:SIGNal
:TCONfigure
:DIRection <name>
:DIRection?
:ASYNchronous
:ILINe <NRf>
:ILINe?
:OLINe <NRf>
:OLINe?
:TRIGger[:SEQuence[1]]
:IMMediate
:COUNt <n>
:AUTO <b>
:AUTO?
:COUNt?
:DELay <n>
:DELay?
:SOURce <name>
:SOURce?
:TIMer <n>
:TIMer?
:SIGNal
Query control source.
Set timer interval (0 to
99999.999 sec).
Query timer.
Loop around control source.
Path to configure Triggers:
Enable (SOURce) or disable
(ACCeptor) Bypass.
Query direction.
Path to configure asynchro-
nous Trigger Link:
Select input line (1 to 6).
Query input line.
Select output line (1 to 6).
Query output line.
Path to program trigger layer:
Immediately perform a scan
step.
Program measure count (1 to
9999, or INF).
Enable (1 or ON) or disable
(0 or OFF) auto-count
(scan-list-length).
Query auto-count.
Query measure count.
Program delay (0 to 99999.999
sec).
Query delay.
Select control source: HOLD,
IMMediate, TIMer, MAN-
ual, BUS, TLINk, EXTernal.
Query control source.
Set timer interval (0 to
99999.999 sec).
Request the programmed timer
interval.
Loop around control source.
Table 16. Trigger command summary (cont.)