10
To recall a channel pattern from memory:
1. Press RECALL and enter the memory location of the
stored channel pattern.
2. Press ENTER. The channel status display will imme-
diately update to reflect the recalled channel pat-
tern.
Another way to recall a channel pattern is to include a
memory location (e.g. M36) in a Channel List and/or
Scan List. In a Channel List, when CLOSE is pressed, the
closed channels in the specified channel pattern will
close, and when OPEN is pressed, the channels will
open. In a Scan List, when the memory location is
scanned, the channel status display will update to
reflect the specified channel pattern. The closed chan-
nels in the channel pattern will remain closed unless a
different channel pattern is scanned.
IEEE-488 bus operation
Over the IEEE-488 bus, the :SAVe and :RECall com-
mands are used to save (store) and recall channel pat-
terns. The following examples demonstrate proper
format:
:MEMory:SAVe M36
:MEMory:RECall M36
As in front panel operation, a memory location can be
included in the Channel List and/or Scan List.
Summary of Contents for 7001