3-16
Remote Operation
Model 6517A Getting Started Manual
Test sequences
:TSEQuence
TSEQuence
subsystem.
:ARM
Arms selected test sequence.
:ABORt
Stops a test sequence in process.
:TYPE <name>
Specify test; DLEakage, CLEakage, CIResistance,
RVCoefficient, SRESistivity, VRESistivity,
SIResistivity, SQSWeep, STSWeep.
:TSOurce <name>
Specify trigger source; MANual, IMMediate, BUS,
TLINk, EXTernal, LCLosure.
:TLINe <NRf>
Specify TLINK line; 1 to 6.
DLEakage
Diode leakage test path:
:STARt <NRf>
Specify start voltage; -1000 to 1000.
:STOP <NRf>
Specify stop voltage; -1000 to 1000.
:STEP <NRf>
Specify step voltage; -1000 to 1000.
:MDELay <NRf>
Specify measure delay; 0 to 99999.9 (sec.).
:CLEakage
Capacitor leakage test path:
:SVOLtage <NRf>
Specify bias voltage (-1000 to 1000).
:STIME <NRf>
Specify bias time; 0 to 99999.9 (sec.).
:DTIMe <NRf>
Specify discharge time; 0 to 9999.9 (sec.).
:CIResistance
Cable insulation resistance test path:
:SVOLtage <NRf>
Specify bias voltage; -1000 to 1000.
:STIME <NRf>
Specify bias time; 0 to 99999.9 (sec.).
:RVCoefficient
Resistor voltage coefficient test path:
:SVOLtage[1] <NRf>
Specify source voltage 1; -1000 to 1000.
:MDELay[1] <NRf>
Specify measure delay 1; 0 to 99999.0 (sec.).
:SVOLtage2 <NRf>
Specify source voltage 2; -1000 to 1000.
:MDELay2 <NRf>
Specify measure delay 2; 0 to 99999.9 (sec.).
:SRESistivity
Surface resistivity test path:
:PDTime <NRf>
Specify pre-discharge time; 0 to 99999.9 (sec.).
:SVOLtage <NRf>
Specify bias voltage; -1000 to 1000.
:STIME <NRf>
Specify bias time; 0 to 99999.9 (sec.).
:MVOLtage <NRf>
Specify measure voltage; -1000 to 1000.
:MTIMe <NRf>
Specify measure time; 0 to 99999.9 (sec.).
:DTIMe <NRf>
Specify discharge time; 0 to 99999.9 (sec.).
:VRESistivity
Volume resistivity test path:
:PDTime <NRf>
Specify pre-discharge time; 0 to 99999.9 (sec.).
:SVOLtage <NRf>
Specify bias voltage; -1000 to 1000.
:STIME <NRf>
Specify bias time; 0 to 99999.9 (sec.).
:MVOLtage <NRf>
Specify measure voltage; -1000 to 1000.
:MTIMe <NRf>
Specify measure time; 0 to 99999.9 (sec.).
:DTIMe <NRf>
Specify discharge time; 0 to 99999.9 (sec.)