2-14
Connections
Model 6485 and 6487 User’s Manual
With Model 6487 front panel operation, an open interlock will display "CLOSE INTLCK"
as an error message when attempting to operate the voltage source on the 50V and 500V
ranges. The Model 6487 in the Model 487 DDC emulation mode displays "IDDCO
ERROR" on the front panel when an "O1" command is sent. The Model 487 displays
"INTERLOCK" for the same condition. The "U9" voltage source error status word func-
tions the same for either the Model 487 or Model 6487 in DDC emulation mode.
Analog output
The Model 6485/6487 has an analog output on the rear panel. The ANALOG OUT pro-
vides a scaled, inverting ±2V output. A full-scale reading corresponds to ±2V output.
CAUTION
Connecting COMMON or ANALOG OUT to earth while floating the
input may damage the instrument.
Connections for using this output are shown in
(Model 6487). For a full-scale input (i.e. 2mA on the 2mA range), the output will be -2V.
The 2V analog output signal is not corrected during calibration. The Model 6485 output
impedance is 1k
Ω
, while the Model 6487 output impedance is <100
Ω
. To minimize the
effects of loading, the input impedance of the device connected to the ANALOG OUT
should be as high as possible.
Figure 2-14
Typical Model 6485 analog output connections
120
FUSE
LINE
630mA
LINE RATING
50, 60Hz
30 VA
T
(SB)
100 VAC
120 VAC
315mAT
(SB)
220 VAC
240 VAC
!
INPUT
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
CAT
I
TRIGGER LINK
RS-232
MADE IN
U.S.A.
220V PK
ANALOG OUT
Test Lead
Model 6485 Rear Panel
Measuring Device
(i.e. Chart recorder)
LO
HI
Summary of Contents for 6485
Page 40: ...1 28 Getting Started Model 6485 and 6487 User s Manual...
Page 94: ...7 8 Remote Operation and Commands Model 6485 and 6487 User s Manual...
Page 95: ...A Specifications...
Page 98: ...A 4 Specifications Model 6485 and 6487 User s Manual...
Page 105: ...C ExamplePrograms Programming examples lists and defines three types of programming exam ples...
Page 113: ......
Page 114: ......