6482-901-01 Rev. A / August 2012
Return to
1-15
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 1: Getting Started
Menus
The following paragraphs discuss the main menu, configuration menus, and rules to navigate
menus.
Main menu
Use the MENU key to access the main menu to select, configure, and/or perform various
instrument operations. These include default setup conditions, communications (GPIB or RS-232),
calibration, front-panel tests, auto zero, timestamp, and numeric display format.
The main menu structure is summarized in
. Use the
to check
and change menu options.
Table 1-3
Main menu
Menu item
1
Description
Parameters
SAVESETUP
SAVE
RESTORE
POWERON
BENCH
GPIB
USER-SETUP-NUMBER
RESET
COMMUNICATION
2
GPIB
RS-232
BAUD
BITS
PARITY
TERMINATOR
FLOW-CTRL
Configure setup conditions.
Save present Model 6482 setup to memory location.
Return the Model 6482 to setup saved in memory.
Select the power-on default setup.
Powers-on to BENCH defaults.
Powers-on to GPIB defaults.
Powers-on to user setup.
Returns instrument to BENCH or GPIB defaults.
Select and configure the remote interface.
Select GPIB (IEEE-488 bus), set primary address,
).
Select the RS-232 interface, set parameters.
Select baud rate.
Select number of data bits.
Select parity.
Select terminator.
Select flow control.
0 to 4
0 to 4
0 to 4
See
0 to 30
(Default: 25)
57600, 38400,
19200, 9600,
4800, 2400,
1200, 600, 300
7 or 8
NONE, ODD,
EVEN
CR, CR+LF, LF,
or LF+CR
NONE or
XON/XOFF
Notes:
1
Top-level menu choices are indicated in bold. Indentation identifies each lower submenu level.
2
When the remote operation interface selection (GPIB or RS-232) is changed, the Model 6482 performs a power-on reset.
To check or change options of the selected interface, you must then re-enter the menu structure.
3
A password is required to unlock calibration. See the
section.
4
Press the
EXIT
key to cancel the test.
5
Disabling auto zero will reduce measurement accuracy.