16-52
Return to
6482-901-01 Rev. A / August 2012
Section 16: SCPI Command Reference
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
DISable <list>
:STATus:QUEue:DISable <list> Disable messages for error queue
Parameters
<list> = (numlist)
where numlist is a specified list of messages that you wish to disable for the error
queue.
Query
:DISable?
Query list of disabled messages
Description
On power-up, all error messages are enabled and will go into the error queue as
they occur. Status messages are not enabled and will not go into the queue. This
command is used to specify which messages you want disabled. Disabled
messages are prevented from going into the error queue.
SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in
Default conditions
PRESet
:SYSTem:PRESet
Return to :SYSTem:PRESet defaults
Description
This command returns the instrument to states optimized for front-panel
operation. :SYSTem:PRESet defaults are listed in the SCPI tables (
through
POSetup
:SYSTem:POSetup <name>
Program power-on defaults
Parameters
<name> = RST
Power-up to *RST defaults
PRESet
Power-up to :SYSTem:PRESet defaults
SAV0
Power-up to setup stored at memory location 0
SAV1
Power-up to setup stored at memory location 1
SAV2
Power-up to setup stored at memory location 2
SAV3
Power-up to setup stored at memory location 3
SAV4
Power-up to setup stored at memory location 4
Query
:POSetup?
Query power-on setup
Description
This command is used to select the power-on defaults. With RST selected, the
instrument powers up to the *RST default conditions. With PRES selected, the
instrument powers up to the :SYStem:PRESet default conditions. Default
conditions are listed in the SCPI tables (
through
With the SAV0-4 parameters specified, the instrument powers-on to the setup that
is saved in the specified location using the *SAV command.