![Keithley 6482 Reference Manual Download Page 210](http://html1.mh-extra.com/html/keithley/6482/6482_reference-manual_4059393210.webp)
6482-901-01 Rev. A / August 2012
Return to
16-17
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 16: SCPI Command Reference
CALCulate[1] and CALCulate2
The CALC1 and CALC2 commands control math functions for channel 1 and channel 2
respectively. They perform the remote equivalents to the front panel CONFIG MSR1 and CONFIG
MSR2 configuration menus.
Select math function
FORMat <name>
:CALCulate[1]:FORMat <name> Select channel 1 math function
Parameters
<name> = MXB[1]
MX + B
COND[1]
I1/V1
POWER[1]
V1*I1
RES[1]
V1/I1
Query
:FORMat?
Query selected math function
Description
This command selects the channel 1 math function. The channel 1 current
reading (I1) and source voltage (V1) are used to calculate COND1 (conductance),
POWER1, and RES1 (resistance), while the MX + B parameter is programmed
separately (see below).
FORMat <name>
:CALCulate2:FORMat <name> Select channel 2 math function
Parameters
<name> = MXB2
MX + B
COND2
I2/V2
POWER2
V2*I2
RES2
V2/I2
Query
:FORMat?
Query selected math function
Description
This command selects the channel 2 math function. The channel 2 current
reading (I2) and source voltage (V2) are used to calculate COND2 (conductance),
POWER2, and RES2 (resistance), while the MX + B parameter is programmed
separately (see below).
Set MX + B parameters
MBFactor <n>
:CALCulate[1]:KMATh:MBFactor <n>Select channel 1 B (offset) value
:CALCulate2:KMATh:MBFactor <n>Select channel 2 B (offset) value
Parameters
<n> =
-9.99999e20 to Specify B (offset) value for MX + B
+9.99999e20
MINimum
-9.99999e20
MAXimum
+9.99999e20
DEFault
0
Query
:MBFactor?
Query B (offset) value for MX + B
Description
These commands program the B (offset) value for the MX + B math function (see
above). Use CALC1 for channel 1 or CALC2 for channel 2.