External triggering example
In a simple test system, you may want to close a switching channel and then measure the current
of the device under test (DUT) connected to that channel. This test system is shown in
,
which uses a Model 6482 to measure 10 DUTs switched by a Model 7011 multiplexer card in a
Model 7001/7002 Switch System.
Figure 8-5
DUT test system
The trigger link connections for this test system are shown in
. The trigger link of the
Model 6482 is connected to the trigger link (IN or OUT) of the switching mainframe. Note that with
the default trigger settings of the switching mainframe, line 1 is an input, and line 2 is an output.
For this example, the Model 6482 and switching system (mainframe) are configured as follows.
Model 6482 setup
Step 1. Restore bench defaults.
1. Press the
MENU
key, select
SAVESETUP
, then press the
ENTER
key.
2. From the SAVESETUP menu, select
RESET
, then press the
ENTER
key.
3. Select
BENCH
, and then press the
ENTER
key.
Step 2. Configure channel.
1. Press the
MSR1
or
MSR2
key to select the channel to measure, and then select the range.
2. Press
SRC1
or
SRC2
to select the voltage source, and then set the source to the value using
the EDIT or numeric keys.
Step 3. Set up trigger parameters.
1. Press the
CONFIG
key, and then the
TRIG
key to access the CONFIGURE TRIGGER menu.
2. Select
TRIG-LAYER
and then press the
ENTER
key to access the CONFIGURE TRIG-LAYER
menu.
Step 4. Set trigger-in event to TRIGGER-LINK.
1. Select
TRIGGER-IN
and then press the
ENTER
key.
2. Select
TRIGGER-LINK
and then press the
ENTER
key.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
SERIAL NO.
120
LINE RATING
50, 60Hz
60 VA MAX
FUSE
LINE
630 mAT
(SB)
315 mAT
(SB)
100 VAC
120 VAC
220 VAC
240 VAC
CAT
I
MADE IN
U.S.A.
OUTPUT ENABLE
RS-232
TRIGGER LINK
INPUT
RATINGS MAX.
30 V @ 20mA
CHANNEL 1
CHANNEL 2
RATINGS MAX.
COMMON
MODE
200V
!
!
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
ANALOG
OUT
RATINGS MAX.
30 V @ 20mA
VOLTAGE
SOURCE
OUT
30 V @ 20mA
HI
1
DUT
#1
2
DUT
#2
10
DUT
#10
Card 1
Model 6482
Triaxial cable
7011 Multiplexer card