GPIB 488.1 Protocol
G-3
Protocol differences
The following information covers the differences between the 488.1 protocol and the SCPI
protocol.
Message exchange protocol (MEP)
When the 488.1 protocol is selected, the MEP is disabled to speed up GPIB operation.
The following guidelines/limitations must be followed when using the 488.1 protocol:
•
If a query is sent, it must be the only command on the line (this limitation also means
no multiple queries can be sent). Otherwise, full SCPI command syntax is still sup-
ported including long-form and short form commands, multiple commands, and MIN/
MAX/DEF parameter de
fi
nitions.
For example, the following command strings are invalid:
:VOLT:RANG 10;*OPC?
:RES:RANG?;:READ?
:READ?;:READ?
The following command strings are valid:
:SOUR1:VOLTage:STARt 1;STOP 10;step 1
:volt:nplc 1.0;:curr:rang min
:RES:RANG? MAX
:READ?
•
When a query is sent, either the data must be read back or a Device Clear (DCL) or
Interface Clear (IFC) must be performed to reset the query.
•
When sending a command or query, do not attempt to read data from the Model 6430
until the terminator has been sent (usually Line Feed with EOI). Otherwise, a DCL or
IFC must be sent to reset the input parser.
•
When receiving data, all data, up to and including the terminator (LF with EOI), must
be accepted. Otherwise, a DCL or IFC must be sent to reset the output task.
•
Empty command strings (terminator only) should not be sent.
Using SCPI-based programs
In general, an existing SCPI-based program will run properly and faster in the 488.1 proto-
col as long as it meets the above guidelines and limitations.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......