Accuracy calculations
The following information discusses how to calculate accuracy for both sense and source
functions.
Measure accuracy
Measurement accuracy is calculated as follows:
Accuracy = ±(% of r offset)
As an example of how to calculate the actual reading limits, assume that you are measuring
10V on the 20V range. You can compute the reading limit range from one-year measure voltage
accuracy specifications as follows:
Accuracy = ±(% of r offset)
= ±[(0.015%
×
10V) + 1.5mV]
= ±(1.5mV + 1.5mV)
= ±3mV
Thus, the actual reading range is 10V ±3mV or from 9.997 to 10.003V.
DC current measurement calculations are performed in exactly the same manner using the
pertinent specifications, ranges, and input signal values.
Source accuracy
Source accuracy is calculated similarly, except source specifications are used. As an exam-
ple of how to calculate the actual source output limits, assume that you are sourcing 0.7mA on
the 1mA source range. You can compute the reading limit range from source current one-year
accuracy specifications as follows:
Accuracy = ±(0.034% of 200nA offset)
= ±
[(0.034%
×
0.7µ
Α
) + 200
νΑ
)]
= ±(238nA + 200nA)
= ±438nA
In this case, the actual current output range is 0.7mA ±438nA or from 0.69956mA to
0.70044mA.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......