Digital I/O Port, Interlock, and Output Configuration
12-9
GUARD
With this output-off state, the current source is selected and set to 0A. Voltage compliance is
set to 0.5% full scale of the present voltage range. This output-off state should be used when
performing 6-wire guarded ohms measurements or for any other load that uses an active
source.
NOTE
When changing the output-off state with the output off, the selected output-off state
will be entered immediately.
Output off states and inductive loads
The output off state you select for inductive loads depends on how much energy the inductor
holds. The Normal output off state is
not
recommended as it lowers the compliance setting. The
Zero or the Guard state are better suited, as Zero does
not
change the compliance setting and
the Guard output off state would change the voltage source to a current source with a voltage
compliance. The Guard state is typically used only for guarded ohms measurements.
To protect the unit from inductive energy, the application may require a spark gap across the
INPUT HI and LO terminals. The SourceMeter does not have internal spark gap protection, as
some leakage current (nA) is associated with the protection circuits.
Remote output con
fi
guration
Output con
fi
guration commands
Table 12-2 summarizes output con
fi
guration commands. These commands include those to
enable and disable the interlock as well as commands to control output off states. See Section
17,
OUTPut subsystem
and
SOURce subsystem
for more information.
Table 12-2
Output configuration commands
Command
Description
:OUTPut:INTerlock:STATe <state>
:OUTPut:INTerlock:TRIPped?
:OUTPut:SMODe <name>
:SOURce:CLEar
:SOURce:CLEar:AUTO <state>
:SOURce:CLEar:AUTO:MODE <name>
Enable/disable interlock (state = ON or OFF).
Query interlock tripped state (1 = tripped).
Select output-off mode (state = NORMal, ZERO, or GUARd).
Turn output source off when in idle state.
Enable/disable auto output-off. State = ON (output off after
measurement) or ON (output stays on).
Auto clear mode. Name = ALWays (every reading; default) or
TCOunt (ON when trigger layer entered; OFF when leaving
trigger layer).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......