12-6
Digital I/O Port, Interlock, and Output Configuration
Safety interlock
The Digital I/O Port provides an interlock line for use with a test
fi
xture interlock switch.
When properly used, the OUTPUT of the SourceMeter will turn OFF when the lid of the test
fi
xture is opened. See
Connections
in Section 2 for important safety information when using
the test
fi
xture interlock.
When the interlock is enabled (see
Front panel output configuration
later in this section), the
output of the SourceMeter cannot be turned on unless the interlock line is pulled low through a
switch to ground as shown in Figure 12-4A. If the lid of the test
fi
xture opens (Figure 12-4B),
the switch opens, and the interlock line goes high turning the OUTPUT of the SourceMeter
OFF (high impedance). The output can only be turned back on by
fi
rst closing the lid of the test
fi
xture and then pressing the OUTPUT ON/OFF key.
NOTE
Interlock can be driven by Digital I/O. Allow 100µs settling and response time. The
Digital I/O lines are open-collector, edge-sensitive and signals should be de-
bounced to avoid erratic operation.
/Interlock
(pin 8)
GND
(pin 5 or 9)
Interlock Switch
(Lid Closed)
Interlock-
Digital I/O
SourceMeter
Test Fixture
A. SourceMeter OUTPUT can be turned on.
/Interlock
(pin 8)
GND
(pin 5 or 9)
Interlock Switch
(Lid Open)
SourceMeter
Test Fixture
B. SourceMeter OUTPUT turns off.
Interlock-
Digital I/O
Figure 12-4
Using test fixture
interlock
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......