Limit Testing
Category register component handler
When using this type of handler, the SourceMeter sends a bit pattern to three handler lines
when a pass or fail condition occurs. This bit pattern determines the bin assignment for the
DUT. With the pass/fail pattern on the output, line #4 is then pulsed. This EOT (end-of-test)
pulse latches the bit pattern into the register of the handler, which places the DUT in the
assigned bin. When interfacing to this type of handler, a maximum of eight component handler
bins are supported.
If the handler requires a high-going or low-going EOT pulse, program SourceMeter for 3-bit
operation and appropriate EOT mode.
NOTE
The EOT and 3-bit modes are configured from the CONFIG LIMIT MENU. See
“Configuring limit tests” later in this section.
Basic binning systems
Two basic binning systems are shown in Figures 11-7 and 11-8. Both systems require a han-
dler to physically place the device packages in the appropriate bins. The handler is controlled
by the SourceMeter via the Digital I/O port.
Single-element device binning
Figure 11-7 shows a basic binning system for single-element devices (i.e., resistors). After
all programmed testing on the DUT is completed, the appropriate digital output information is
sent to the component handler, which then places the DUT in the appropriate bin. The compo-
nent handler selects the next DUT, and the testing process is repeated.
Handler
Dig
In
DUT
HI
LO
IN/OUT
6430
Dig
I/O
Figure 11-7
Binning system - single
element devices
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......