Model 6220/6221 Reference Manual
Output Connections
2-5
Return to
Output configurations
There are four configurations that can be used for current source output:
•
Triax inner shield is connected to
Output Low
, which is connected to
Earth
Ground
(see
•
Triax inner shield is connected to
Cable Guard
and
Output Low
is con-
nected to
Earth Ground
A).
•
Triax inner shield is connected to
Output Low
, which is disconnected from
Earth Ground (
Floating
) (see
B). This is the default setting.
•
Triax inner shield is connected to
Cable Guard
.
Output Low
is accessed at
the LO banana jack and is disconnected from Earth Ground (
Floating
) (see
Using the Model 622x with the triax inner shield connected to Output Low makes it
compatible with the Keithley Model 220 current source.
WARNING
To prevent electric shock and/or damage to the Model 622x, DO
NOT exceed the maximum (Max) voltage levels indicated in
.
CAUTION
To prevent damage to internal fuses of the Model 622x, current
from output low to earth ground must be limited to 1A. Use a
fuse or other current limiting device in the external test circuit.
One internal fuse is located between the Triax Output Low and
banana jack LO. The other internal fuse is located between
Triax Output Low and earth ground.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...