Model 6220/6221 User’s Manual
Sweeps
4-11
Return to
SCPI commands — sweeps
Commands for linear and logarithmic staircase sweeps are listed in
,
while commands for custom (list) sweeps are listed in
Table 4-3
Staircase sweep commands (linear and logarithmic)
Command
Description
Default
SOURce[1]:CURRent:STARt <n>
Sets start current.
<n> = -0.105 to 0.105 (A)
0
SOURce[1]:CURRent:STOP <n>
Sets stop current.
<n> = -0.105 to 0.105 (A)
0.1
SOURce[1]:CURRent:STEP <n>
Sets step current.
<n> = 1e-13 to 0.105 (A)
1e-2
SOURce[1]:CURRent:CENTer <n>
Sets center current.
<n> = -0.105 to 0.105 (A)
0
SOURce[1]:CURRent:SPAN <n>
Sets span current.
<n> = 2e-13 to 0.210 (A)
0.1
SOURce[1]:DELay <n>
Sets source delay.
<n> = 0.001 to 999999.999 (s)
1
SOURce[1]:SWEep:ARM
Prepare for running the sweep.
SOURce[1]:SWEep:ABORt
Abort sweep immediately.
SOURce[1]:SWEep:SPACing <name>
Selects sweep type.
<name>=LINear, LOGarithmic, or LIST
LINear
SOURce[1]:SWEep:POINts <n>
Sets sweep points.
<n> = 1 to 65535
11
SOURce[1]:SWEep:RANGing <name>
Selects sweep source ranging.
<name>=AUTO, BEST, or FIXed
BEST
SOURce[1]:SWEep:CABort <b>
Sets compliance sweep abort state.
<b> = ON or OFF
OFF
SOURce[1]:SWEep:COUNt <NRf>
Sets sweep count.
<NRf> = 1 to 9999 or INFinite
1
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......