Model 6220/6221 User’s Manual
Delta, Pulse Delta, and Differential Conductance
5-41
Return to
The following query command is used to return the response to a query
command sent over the serial port:
SYSTem:COMMunicate:SERial:ENTer?
When communicating over the serial port, there are no errors reported if a
Model 2182/2182A is not connected to the serial port.
Examples
– The following commands demonstrate proper syntax for
sending commands and returning responses to queries over the serial port:
4.
Set measurement units
– Volts are the default units for the Model 622x
but can instead be expressed (and displayed) as an Ohms, Watts, or Sie-
mens reading (see “
for details). The
commands to set measurement units are listed in
. The following
example shows the command to select conductance (Siemens) measure-
ment units:
5.
Set up, arm, and run Differential Conductance
– Details on the com-
mands to set up and arm Differential Conductance are provided in
.
The following demonstrates the proper command sequence to set up, arm,
and run Differential Conductance for the Example in
:
SYST:COMM:SER:SEND “VOLT:RANG 2”
‘
‘
Select 2V range for 2182/
2182A.
SYST:COMM:SER:SEND “VOLT:RANG?”
‘ Send range query.
SYST:COMM:SER:ENT?
‘ Return response to query.
SYST:COMM:SER:SEND “VOLT:NPLC 1”
‘
‘
Set rate to 1PLC for 2182/
2182A.
SYST:COMM:SER:SEND “VOLT:NPLC?”
‘ Send rate query.
SYST:COMM:SER:ENT?
‘ Return response to query.
UNIT S
‘ Select Siemens measurement units.
*RST
‘ Restores 622x defaults.
SOUR:DCON:STARt 0
‘ Sets start value to 0µA.
SOUR:DCON:STEP 10e-6
‘ Sets step size to 10µA.
SOUR:DCON:STOP 50e-6
‘ Sets stop value to 50µA.
SOUR:DCON:DELTa 20e-6
‘ Sets delta value to 20µA.
SOUR:DCON:DELay 1e-3
‘ Sets delay to 1ms.
SOUR:DCON:CAB ON
‘ Enables Compliance Abort.
TRAC:POIN 6
‘ Sets buffer size to six points.
A
SOUR:DCON:ARM
‘ Arms Differential Conductance.
INIT:IMM
‘ Starts measurements.
B
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......