4210-SMU sink boundaries
Nominal 4210-SMU sink boundaries are shown in the next figure. Actual boundaries are 210 V at 105
mA or 21 V at 1.05 A.
Figure 66: 4210-SMU and 4200-PA operating boundaries
Source-measure considerations
When configured to source current (I-Source), the SMU functions as a high-impedance current
source with voltage limit capability that can measure current (I-Meter) or voltage (V-Meter). The