Appendix C: Using a Keysight 4284/4980A LCR Meter
Model 4200A-SCS Parameter Analyzer Reference Manual
C-2
4200A-901-01 Rev. C / February 2017
The following figure shows the three regions of a typical C-V curve for a MOS capacitor.
Figure 570: Typical C-V curve for a MOS capacitor
Capacitance measurement tests
The 4200A-SCS provides the following user modules to perform C-V tests using the Keysight Models
4284A and 4980A:
•
CvSweep4284.
C-V sweep test: Performs a capacitance and conductance measurement at each
step of a user-configured linear voltage sweep.
•
Cmeas4284.
C measurement: Performs a capacitance and conductance measurement at a fixed
bias voltage.
Details on the user modules for the Keysight Models 4284A and 4980A are provided in the
HP4284ulib User Library Reference
(on page 6-309).
If needed, you can initially do an OPEN and SHORT correction on the 4284A or 4980A to achieve
the most accurate C-V measurements. See the Keysight 4284A or 4980A
Operation Manual
for
details.