Model 3761 10-Channel Low Current Multiplexer Card User's Manual 

Section 5: Typical applications 

3761-900-01 Rev. B

 

/

 

January 2018 

5-5

The capacitor insulation resistance is calculated using the following formula: 

Where: 

R

INSL

 = Insulation resistance of the capacitor under test

V

TEST

 = The test voltage

V

MEAS

 = Voltage measurement at the output of the 3761

R

DCHG

 = Value of the discharge resistor

Be sure to select a series resistor value that maintains the 3761 test voltage specifications, 

and select a V

EQ

 value that minimizes the settling time of the indirect measurement. 

Connecting multiple 3761 cards 

Two output connectors on each 3761 card let you add multiple cards to expand the size of the 

multiplexer by ten for each additional 3761. An additional relay isolates all circuitry on the card from 

the outputs when all channels are open and closes when any channel closes. This isolation relay 

configures itself automatically when assembling a test system using multiple 3761 cards. 
The 3706A supports up to six 3761 cards, for a maximum of 60 low-current channels in a single 

mainframe system. Up to sixty-four 3706A systems can be configured to provide 3,840 low-current 

channels. See the 

Series 3706A System Switch/Multimeter Reference Manual

, available on the 

Keithley Instruments website

 (

http://www.tek.com/keithley

), for multiple 3706A system configurations. 

Leakage current specifications increase for each additional card enabled in a test system. 

A 20-channel scanner system is shown in the next figure. In this configuration, you will connect one 

output connector of the 3761 to a picoammeter, and the second output connector to an output 

connector of a second 3761 card. 

Use a SMA-male to SMA-male cable to connect the cards, such as a Cinch Connectors part 

number 415-0038-012. 

Summary of Contents for 3761

Page 1: ...User s Manual 3761 900 01 Rev B January 2018 P3761 900 01B 3761 900 01B www tek com keithley Model 3761 10 Channel Low Current Multiplexer Card...

Page 2: ...s is strictly prohibited These are the original instructions in English All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments Other brand names are tra...

Page 3: ...d to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exerc...

Page 4: ...lains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or m...

Page 5: ...nstallation 2 2 Web browser support 2 3 General operation 3 1 General operation 3 1 Safety hazards 3 1 Current and voltage switching applications 3 2 Operation considerations 4 1 Cables 4 1 External f...

Page 6: ...re 1 Model 3761 10 channel Low Current Multiplexer Card The 3761 has two output connectors which allow you to add additional cards Each additional card increases the capacity of the multiplexer by ten...

Page 7: ...nded warranty Additional years of warranty coverage are available on many products These valuable contracts protect you from unbudgeted service expenses and provide additional years of protection at a...

Page 8: ...http www tek com keithley General information General ratings for the 3761 are listed in the following table Category Specification Voltage rating 30 V Current rating 100 mA Input and output connecti...

Page 9: ...lified personnel Do not perform these procedures unless you are qualified Failure to recognize and observe standard safety precautions could result in personal injury or death due to electric shock Un...

Page 10: ...puts are located on the front of the card Figure 2 Model 3761 card connectors While SMA connectors are not typically used in low current applications the connectors on the 3761 are compact Teflon insu...

Page 11: ...last inch or so press in firmly to mate the card connector to the mainframe connector 5 On each side of the card there is a mounting screw Tighten these two screws to secure the card to the mainframe...

Page 12: ...and any unprotected leads wiring with double insulation for 500 volts Category I Follow these recommendations before and after applying power to the equipment 1 Do not exceed the 3761 maximum allowab...

Page 13: ...cations For current switching applications each channel on the 3761 has a pair of single pole single throw SPST normally open NO relays The relay pair switches signal HI between signal LO analog groun...

Page 14: ...a low impedance picoammeter When a channel is not selected for measurement the current path is maintained through that channel s analog grounding relay which is automatically updated with the channel...

Page 15: ...voltsampsmode to any value other than 0 or 1 will result in the following error 222 Parameter data out of range slot slot_3761 voltsampsmode 2 setting voltsampsmode on a card that does not support it...

Page 16: ...apacitance Charged capacitances in the circuit can cause excessive currents surge currents to flow through the relay contacts if the source is shorted out or if the output is switched to capacitive lo...

Page 17: ...nt Multiplexer Card User s Manual 4 2 3761 900 01 Rev B January 2018 High impedance measurements Because of the high impedance circuits on the 3761 this card contains no user serviceable components an...

Page 18: ...f the scanner card include Low offset current error 1 pA specified 30 fA typical Switching of AC or DC signals up to 100 mA SMA input and output connectors Two outputs for connecting multiple 3761 car...

Page 19: ...h up to 1 MHz Maximum switched and carry current and voltage levels are 100 mA and 30 V peak using resistive loads Direct measurement of capacitor leakage or insulation resistance If you want to measu...

Page 20: ...citor under test 3 When the capacitor is charged close the applicable channel of the 3761 to start the measurement The current is measured at IM as shown in the previous figure To maintain the voltage...

Page 21: ...esistance and an individual discharge resistor The discharge resistors are used to drain the electric charge from the capacitors when the test completes Using the indirect method the Model 3762 can be...

Page 22: ...rd from the outputs when all channels are open and closes when any channel closes This isolation relay configures itself automatically when assembling a test system using multiple 3761 cards The 3706A...

Page 23: ...Section 5 Typical applications Model 3761 10 Channel Low Current Multiplexer Card User s Manual 5 6 3761 900 01 Rev B January 2018 Figure 8 20 channel scanner system with two 3761 cards in ampere mode...

Page 24: ...xternal factors 4 1 G General information 1 3 General operation 3 1 H Handling precautions 2 1 High impedance measurements 4 2 I Indirect measurement of capacitor leakage or insulation resistance 5 2...

Page 25: ...names are the property of Keithley Instruments All other trademarks and trade names are the property of their respective companies Keithley Instruments Corporate Headquarters 28775 Aurora Road Clevel...

Reviews: