Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-205
dmm.measure()
This function returns the last reading of the measurement process without using the trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
reading
= dmm.measure()
reading
= dmm.measure(
bufferVar
)
reading
The last reading of the measurement process
bufferVar
A previously created reading buffer where all readings are stored
Details
This is not available for
"nofunction"
. If the command is queried when
"nofunction"
is
selected,
nil
is returned. If it is written when
"nofunction"
is selected, an error is returned.
When you use a reading buffer with a command or action that makes multiple readings, all readings
are available in the reading buffer. However, only the last reading is returned as a reading with the
command.
If you define a specific reading buffer, the reading buffer must exist before you make the
measurement.
You can also use a reading buffer to store additional information that is acquired while making a
measurement.
The
dmm.measurecount
attribute determines how many measurements are performed. When you
use a buffer, it also determines if the reading buffer has enough room to store the requested
readings. The amount of room is based on readings already stored in the buffer (
bufferVar
.n
), the
capacity of the buffer (
bufferVar
.capacity
), and the append mode of the reading buffer
(
bufferVar
.appendmode
). If the append mode is set to 0, any stored readings in the buffer are
cleared before new ones are stored. If append mode is set to 1, any stored readings remain in the
buffer and new ones are added to the buffer after the stored ones.
Example
DCVBuffer = dmm.makebuffer(100)
dmm.func = "dcvolts"
dmm.measurecount = 100
dmm.measure(DCVBuffer)
Performs 100 DC voltage measurements and
stores them in a buffer called
DCVBuffer
.
Also see
(on page 11-16)
(on page 11-19)
(on page 11-28)
(on page 11-207)