Section 1: Verification procedures 3390 50 kHz Arbitrary Waveform / Function Generator Verification and Adjustment Manual
1-2
3390-905-01Rev. B / May 2016
Recommended test equipment
The table below identifies the recommended equipment needed for completing the verification and
adjustment procedures. In the case that the instrument listed is not available, you may substitute it for
an instrument with similar specifications.
Recommended
equipment
Description
Requirements
Keysight 3458A
8-1/2 Digit Multimeter
AC volts, true RMS, AC coupled, accuracy: ±0.02 % to
1 MHz
DC volts, accuracy: 50 ppm, resolution: 100
μV
Resistance, offset-compensated, accuracy: ±0.1
Ω
Keysight 53132A
Universal counter
Accuracy: 0.1 ppm
Keysight E4418B
Power meter
100 KHz to 100 MHz
1
μW to 100 mW (–30 dBm to +20 dBm)
Accuracy: 0.02 dB, resolution: 0.01 dB
Keysight 8482A
Power sensor
100 KHz to 100 MHz
1
μW to 100 mW (–30 dBm to +20 dBm)
Cable
BNC to BNC cable
2 set
Cable
T-type BNC connector
(female/male/female)
1 set
Cable
BNC female to double
stacking banana plug
1 set
Cable
Adapter-coaxial straight
female-bnc male-n 50
Ω
1 set
Cable
Adapter-coaxial straight
male-bnc female-n 50
Ω
1 set
Optimal test conditions
All procedures should comply with the following conditions for optimum performance:
Assure that the calibration ambient temperature is stable and between 21 °C and 25 °C
(23 °C ±2 °C).
•
Assure ambient relative humidity is less than 80 %.
•
Allow a 1-hour warm-up period before verification or adjustment.
•
The measurement cable length needs to be as short as possible and consistent with the
impedance requirements.
•
Only use a 50
Ω cable.