Model 2790 SourceMeter
®
Switch System Reference Manual
2-17
Remote programming — Multiple channel control commands
The commands to close and open the system channel are listed in
.
Reference:
a.
ROUTe:MULTiple:CLOSe <clist>
This command functions like the front panel CLOSE key (MULTI menu option) to
close channels. When you send this command to close the channels specified in the
<clist>, only those listed channels will close. Channels not specified are not
affected, and channel pairing is disabled for the 7702 module.
klqb
Channels closed by ROUT:MULT:CLOS are not displayed.
b.
ROUTe:MULTiple:OPEN <clist>
With this command, you can open one or more switching module channels. When
you send this command to open the channels specified in the <clist>, only those
listed channels will open. Channels not specified are not affected.
c.
ROUTe:OPEN:ALL
This command functions the same as the front panel OPEN key (ALL menu
option). It simply opens all channels (including non-measurement channels) in the
mainframe.
Table 2-2
Multiple channel control commands
Commands
Description
Ref
ROUTe:MULTiple:CLOSe <clist>
Specify one or more channels to close.
a
ROUTe:MULTiple:OPEN <clist>
Open channels specified in list. Unlisted
channels not affected.
b
ROUTe:OPEN:ALL
Open all channels.
c
ROUTe:MULTiple:CLOSe?
Returns a <clist> of all closed channels.
d
ROUTe:MULTiple:CLOSe:STATe?
<clist>
Query closed channels in specified list
(1 = closed).
e
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...